System suitability tests (SST) confirm the chromatographic system is fit for use on the day.
Repeatability. Six replicate standard injections give areas of 1,250,000, 1,262,000, 1,245,000, 1,258,000, 1,240,000 and 1,255,000. Mean = 1,251,667; standard deviation = 8,262; %RSD = 0.66%. Assay methods commonly require an RSD of not more than about 2.0% for replicate standards.
Column efficiency. With a retention time of 6.0 min and a peak width at the base of 0.30 min, the theoretical plate number N = 16 x (tR / W)^2 = 16 x 20^2 = 6,400.
Peak symmetry. The tailing factor T = W0.05 / (2f), where W0.05 is the peak width at 5% of peak height and f is the distance from the leading edge to the peak maximum at that height. With W0.05 = 0.24 min and f = 0.10 min, T = 1.2. Many methods require T of not more than 2.0.
| SST parameter | What it shows | Common requirement |
|---|
| %RSD of replicate injections | Precision of the system | Not more than about 2.0% for assay |
| Theoretical plates (N) | Column efficiency | Method-specific minimum |
| Tailing factor (T) | Peak symmetry | Not more than about 2.0 |
| Resolution (Rs) | Separation of critical peaks | Often at least 1.5 to 2.0 |
The monograph or validated method always sets the actual limits.